Maryam Saeedullah, Syed Wilayat Husain.
Comparison of Geometric Design of a Brand of Stainless Steel K-Files: An In Vitro Study.
J Coll Physicians Surg Pak Jan ;28(4):327-9.

The purpose of this experimental study was to determine the diametric variations of a brand of handheld stainless-steel K-files, acquired from different countries, in accordance with the available standards. 20 Mani stainless-steel K-files of identical size (ISO#25) were acquired from Pakistan and were designated as Group A while 20 Mani K-files were purchased from London, UK and designated as Group B. Files were assessed using profile projector Nikon B 24V. Data was statistically compared with ISO 3630:1 and ADA 101 by one sample T test. Significant difference was found between Groups A and B. Average discrepancy of Group A fell within the tolerance limit while that of Group B exceeded the limit. Findings in this study call attention towards adherence to the dimensional standards of stainless-steel endodontic files.